Vögtlin red-y smart Tryckregulator

Fakta
0
Artikelnummer

GSP-Vögtlin

Smart tryckregulator

Elektronisk tryckregulator, smart pressure controller, för gaser med integrerad flödesmätning.

Beskrivning:

  • Reglerar mot flöde eller tryck

  • Ned till 0,3% FS

  • 3 års garanti

  • Extremt snabba och noggranna mätningar

Teknisk data
1
Red-y smart tryckregulator

Teknisk Översikt

Instrument versions flow Standard Accuracy: ± 1.0 % of full scale
Turndown ratio: 1 : 50
Hi-Performance Accuracy: ± 0.3 % of full scale + ± 0.5% of reading
Turndown ratio: 1 : 100
Instrument versions pressure Pressure control Accuracy: ± 0.5 % of full scale
Back pressure control Accuracy: ± 0.5 % of full scale
Differential pressure controller according to customer specifications
Performance data Media (real gas calibration) Air, O2, N2, He, Ar, CO2, H2, CH4, C3H8
Other gases and gas mixtures on request
Response time 50 ms
Repeatability ± 0,2% of full scale
Longterm stability < 1% of measured value / year
Power supply 24 Vdc (18 – 30 Vdc), 15 Vdc on request
Current consumption max. 250mA
Temperature 0 – 50°C
Materials Anodized aluminium, optional stainless steel electropolished
Seals FKM, optional EPDM
Pressure sensitivity < 0.2% / bar of reading (typical N2)
Temperature sensitivity < 0.025% FS measuring range type / °C
Measuring ranges
Flow (Air) Type Measuring range (Air), full scale freely selectable
pressure GSP GSX-A from 0 ... 25 mln/min to 0 ... 600 mln/min
back pressure GSB GSX-B from 0 ... 600 mln/min to 0 ... 6000 mln/min
GSX-C from 0 ... 6 ln/min to 0 ... 60 ln/min
GSX-D from 0 ... 60 ln/min to 0 ... 450 ln/min
Pressure Full scale gauge pressure 0.5 bar ü, 1 bar ü, 2 bar ü, 5 bar ü, 10 bar ü
Full scale absolute pressure 1.2 bar a, 2 bar a, 5 bar a, 10 bar a
Integration Output signals:
- Analog 0..20 mA, 4..20 mA, 0..5 V, 1..5 V, 0..10 V, 2..10 V
(for actual value flow only)
- Digital RS-485; Modbus RTU (Slave); Lab View-VI‘s available
Option: ProfiBus DP-V0, DP-V1
(for pressure & flow)
Process connection G¼¨ female less than 60 ln/min, G½¨ female less than 450 ln/min
Inlet section non required
Electrical connection Sub D plug, 9 pole
Mounting orientation Any orientation (horizontal only above 5 bar)
Safety Test pressure 16 bar a
Leak rate < 1 x 10-6 mbar l/s He
Environmental protection IP-50
EMC EN 61326-1
Priser
1
Priser/Offert

Välkommen att kontakta oss för information om priser och övriga frågor. Fyll i formuläret, offertförfrågan, nedan så hör vi av oss inom kort. Du kan också kontakta oss via:

Telefon: +46 (0)31 748 17 80
Fax: +46 (0)31 748 17 85
Mail: info@labinett.se

Obs: Teknisk support ingår i våra priser

Relaterade Produkter
1
Fler Produkter
red-y smart back pressure controller GSB Red-y smart back pressure controller GSB
Electronic back pressure controller with integrated flow measurement

För fler relaterade produkter, se här